Table 3 Accuracy of reproducing the intraoperative validated angle as assessed by the post-operative radiographs1. Data are presented as means ± standard deviation (95% confidence intervals)

From: Cup accuracy and early-term clinical outcomes of a novel, pinless, robotic-assisted total hip arthroplasty system: A first-in-human pilot study

 

Intra-operative Validated Angle

Post-operative Radiographs

P-value

Mean Absolute Error

Anteversion

19.7° ± 3.0° (18.2° – 21.1°)

21.0° ± 3.9° (19.04° – 22.9°)

0.241

3.1° ± 2.5°

(1.9° – 4.4°)

Inclination

40.6° ± 2.2° (39.5° – 41.6°)

43.56° ± 4.13° (41.5° – 45.6°)

 < 0.001

3.9° ± 2.2°

(2.8° – 5.0°)

  1. Validated using the robotic software