Skip to main content

Table 5 Summary of achieved cup accuracy from previous published studies using ROSA or MAKO raTHA compared to the current study

From: Cup accuracy and early-term clinical outcomes of a novel, pinless, robotic-assisted total hip arthroplasty system: A first-in-human pilot study

Author

Robotic System

Mean error between target and achieved alignment

Percentage within safe zone

Inclination

Anteversion

Lewinnek

Callanan

Liu (Current Study)

ROSA

3.8° ± 2.6°

2.9° ± 2.4°

88.9%

61.1%

Kamath [17]

ROSA Cadaveric

1.8° ± 1.3°

2.6° ± 2.3°

NR

100%

Buchan [35]

ROSA

2.8° ± 5.2°

3.5° ± 5.1°

81.6%

NR

Ong [34]

ROSA

2.25° ± 5.55°

4.09° ± 6.14°

77%

NR

Redmond [38]

MAKO

3.3° ± 3.1°

2.9° ± 2.3°

NR

NR

Domb [39]

MAKO

Posterior

Posterior

Posterior

Posterior

0.13° ± 3.33°

4.89° ± 3.87°

97.78%

94.07%

Anterior

Anterior

Anterior

Anterior

0.78° ± 4.88°

0.56° ± 4.81°

87.10%

77.42%

Stewart [12]

MAKO

3.80° ± 4.41°

3.64° ± 3.13°

75%

NR