Lewinnek zone not “the be-all and end-all” functional planning for acetabular component positioning in total hip arthroplasty

Abstract

Background

Proper positioning of a total hip arthroplasty (THA) plays a crucial role in the success and long-term survivorship of the implant. Cup positioning within the Lewinnek Safe Zone (LSZ) does not, however, avoid implant dislocation. Thus, the concept of a functional cup position has been introduced. The purpose of this study was to assess the discrepancy between LSZ and the acetabular cup position suggested by the patient’s specific functional planning. The hypothesis was that a mismatch does exist.

Methods

One hundred consecutive patients with primary hip osteoarthritis undergoing primary THA with a personalized functional preoperative planning and patient-specific cup implantation system were enrolled. Anatomical and spino-pelvic functional parameters were recorded and, for each patient, a “safe cup orientation” was suggested. The suggested functional safe zone was compared to the LSZ.

Results

The mean suggested inclination was 39° ± 3° (range 32°–45°). The mean suggested anteversion was 21° ± 3° (range 12°–28°). The patient’s functional acetabular inclination (AI) corresponded to the LSZ in one of the 100 patients, whereas the acetabular anteversion (AV) was outside the LSZ in 8 of the 100 patients. The mean pelvic tilt while standing and sitting were 0.5° ± 7° (range 21°–45°) and −6° ± 16.7° (range −63°–33°), respectively. The mean pelvic incidence was 52° ± 9.7° (range 33°–83°).

Conclusion

When a functional patient’s specific preoperative planning is performed, the LZS does not correspond to the patient’s functional safe zone in about 8% of patients. The concept of a universal safe zone should be revisited, and a functional personalized safe zone may have to be more widely considered.

Introduction

Proper positioning of a total hip arthroplasty (THA) implant is crucial to the success and long-term survivorship of implants [1,2,3,4,5,6,7,8,9,10,11]. Acetabular cup orientation is considered the most relevant factor [12], and the Lewinnek safe zone (LSZ: inclination 40° ± 10° and anteversion 15° ± 10°) [7] has been traditionally advocated for proper acetabular positioning. However, 60% of dislocations occur even though the acetabular cup is within the LSZ [13,14,15]. Thus, the concept of a functional cup position, which considers changes in anteversion and inclination according to the postural position of the pelvis, has been introduced [16]. Sagittal postural balance varies with patients, and multiple lumbar and spino-pelvic conditions may affect it [17].

An acetabular cup placed within the LSZ intraoperatively may become unsafe during daily activities, and dislocations may therefore occur. Several authors suggested preoperative assessment of spinopelvic balance through dynamic standing and sitting radiographs, lumbosacral radiographs, or dynamic simulation of the spinopelvic movements [18,19,20,21]. The Optimise Positioning System (OPS™, Corin Ltd., Cirencester, UK) platform allows for 3D functional analysis of patients’ hip-spine anatomy and mobility with patient-specific implant positioning. The functional analysis consists of a low-dose CT scan and three lateral radiographs in three functional positions: standing, step-up, and flexed-seated. Hip-spine relationship is assessed through a functional computed analysis and the resultant personalized preoperative planning is provided to obtain the “best” implant orientation for stability and impingement-free ROM.

The purpose of this study was to assess the discrepancy between LSZ and the acetabular cup position suggested by OPS patient’s specific functional analysis. The hypothesis was that a mismatch does exist.

Materials and methods

This was an IRB-approved, prospective cohort, observational imaging study. Between October 2020 and March 2021, 100 consecutive patients with primary hip osteoarthritis undergoing primary THA with a personalized dynamic preoperative planning and patient-specific cup implantation system were enrolled in the study (Fig. 1). All procedures were performed by a single fully trained orthopedic hip surgeon.

Fig. 1
figure 1

The flow diagram illustrates patient selection and treatment

Exclusion criteria included: (1) history of previous acetabular/pelvis fracture or hip infections altering the normal bony anatomy, (2) neurological or motor disorders, (3) developmental dysplasia of the hip, and (4) previous spinal surgery.

The imaging study (OPS™, Corin Ltd., Cirencester, UK) consisted of an anteroposterior radiograph of the pelvis, three functional lateral spinopelvic radiographs (standing, flexed-seated and stepping-up), and a low-dose CT scan (mean dose 2.8 to 4.1 mSv per scan) of the lower limbs.

The images were then sent to the manufacturer for analysis.

The anterior and lateral radiographs allowed for the evaluation of the following parameters (Figs. 2 and 3).

  • Pelvic Incidence (PI): the angle between a vertical line through the femoral head and a line from the mid-sacral plateau and the femoral head which denotes the pelvic orientation in space.

  • Sacral Slope (SS): The angle between the sacral plate and the horizontal line. It is reported for the patient’s standing, seated, and step-up positions.

  • Pelvic Tilt (PT): the angle between the segment linking the middle part of the sacral plate with the bicoxofemoral axis and a vertical line crossing the interfemoral axis. PT is positive when the hip is in front of the sacral plate and negative when it is behind it.

  • Lumbar lordosis (LL) is measured as the angle between the upper-end plate of S1 and the upper-end plate of L1.

  • Lumbar flexion (LF) is reported as the change in lumbar lordosis (LL) between the standing and seated positions.

  • PI-LL mismatch: the difference between pelvic incidence (PI) and standing lumbar lordosis (LL).

Fig. 2
figure 2

Lateral radiographs evaluated on OPSinsight with pelvic parameters shown

Fig. 3
figure 3

The OPSInsight general layout provides a range of biomechanical and spinopelvic measurements. The implants used in the plan are presented on the left-hand side of the screen. Here, the cup model, cup size, liner material, stem model, stem size, stem neck, head material, head size, and head offset can be viewed and changed

The differences between hip/spine parameters in different postural positions were calculated in degrees as the variation from the starting position and expressed as ΔPI, ΔPT, ΔSS, etc. From the CT scan the following parameters were evaluated (Fig. 2):

  • Femoral head size (FSH): the measurement in millimeters of the diameter of the femoral head

  • Acetabular size (AS): Acetabular diameter measured in millimeters.

  • Acetabular inclination (AI): the angle between the longitudinal axis and the acetabular axis in the coronal plane.

  • Acetabular version (AV): the complementary angle to the angle between the anterior and posterior wedges of the native acetabulum (or the acetabular cup, postoperatively) and a second line connecting it to the ischiatic spine.

  • Femoral version: the angle between the axis of the native femoral neck and the posterior condylar axis

  • Leg length: the sum of the femur and tibia lengths are taken as the distance between the native femoral head and the knee center and the distance between the knee center and the ankle center, respectively. The measurement is performed on the CT scout with the patient in a lying position.

FHS, AS, AI, and AV were measured on the transverse plane running through the middle of the femoral head at its greatest diameter. As far as I am concerned.

A validated algorithm integrated all these parameters to produce a ROM simulation of the normal and prosthetic hip, suggesting an ideal personalized cup position for each patient. The suggested planning included the contact paths simulation presented as polar plots, a proposed cup type, size, and orientation, and, as far as the femoral component is concerned, the level of the osteotomy, the stem type, position, and size, and the estimated change in the leg length and offset (Figs. 2 and 3). According to his own experience and 3D simulations, the surgeon can adjust all these parameters to create the final plan for the operating theatre. All the patients received an uncemented THA through a direct lateral or direct anterior approach at the discretion of the operating surgeon. As described in our initial validation study for the OPS system, femoral osteotomy and acetabular reaming were performed through PSI guides [21]. 3D printed patient-specific intra-operative instrumentation (provided by Corin Ltd., Cirencester, UK) and a laser-guided technique assisted the senior surgeon (R.I) in placing the prosthetic component in the planned position. All patients received an uncemented acetabular cup (TrinityTM cup, Corin Ltd., Cirencester, UK) and a taper-wedged blade stem (TriFit TSTM, Corin Ltd.).

For each patient, the reported data were collected, and the suggested functional patient-specific safe zone was compared to the LSZ. Descriptive and inferential statistics were performed for demographics and patient-reported outcomes. Continuous variables were summarized using means and standard deviations (SD). A two-sample Student t-test or Wilcoxon rank sum test was used to identify differences.

Results

One-hundred consecutive patients were included in the study (53 [53%] and 47 females [47%]). The mean age was 74 ± 12 (range 59–81) years. The mean BMI was 24.3 (range 23–30). The demographic data are reported in Table 1.

Table 1 Functional preoperative data of patients included in the study

The mean suggested inclination was 39° ± 3° (range 32°–45°). The mean suggested anteversion was 21° ± 3° (range 12°–28°). Regarding AI, the LSZ corresponded to the patient’s specific functional cup position in all patients (100% correspondence). Regarding AV, in 8 patients the functional safe zone was outside the LSZ (92% correspondence).

The mean pelvic tilt while standing and sitting were 0.5° ± 7° (range 21°–45°) and − 6° ± 16.7° (range −63°–33°), respectively. The mean pelvic incidence was 52° ± 9.7° (range 33°–83°). The mean lumbar lordosis while standing and sitting were 58° ± 11.2° (range 24°–83°) and 10° ± 12.9° (range −14°–49°), respectively. The mean lumbar flexion was 47° ± 13.8° (range 10°–76°). The mean pelvic incidence-lumbar lordosis mismatch (PI-LL mismatch) was −6° ± 9.3° (range −24°–19°).

The highest degree of suggested anteversion (28°) occurred in a patient with 11° of posterior pelvic tilt from standing to sitting and a subsequent increased risk of posterior dislocation which, however, did not occur. The lowest degree of suggested anteversion (12°) took place in a patient with a flat back (19° of PI-LL mismatch) and subsequent increased risk of posterior impingement and anterior dislocation. Two complications developed.

One patient suffered an anterior dislocation 4 months after surgery. According to the postoperative CT scan, it might depend on an excessive AV (38° vs planned 24°), a likely result of the acetabular guide malpositioning intraoperatively. The patient underwent THA revision and was then removed from the study. One patient suffered from a periprosthetic joint infection (PJI) and was subsequently excluded. Overall dynamic preoperative data are reported in Fig. 1.

Discussion

The main finding of the present study was that, in 8% of patients, the suggested patient-specific functional cup position was outside the LSZ for the AV. The mean values of the patient-specific functional cup position were 39° ± 3° and 21° ± 3° for AI and AV, respectively. The main difference lay in the suggested cup anteversion, as a markedly higher anteversion (21° ± 3°), as compared to LSF, was found. The highest degree of anteversion (28°) was suggested for a patient with only 11° of posterior pelvic tilt from standing to sitting. The low posterior pelvic tilt reduces the ante-inclination of the cup and may cause an increased risk of posterior dislocation (especially with the posterior approach) during sitting activities [19]. Conversely, the lowest degree of anteversion (12°) was suggested for a patient with 19° of PI-LL mismatch. In case of flat back or high PI, patients presented an increased risk of anterior dislocation, and lower anteversion may be adopted. Regarding cup inclination, the mean value was quite similar (39° ± 3°) to LSZ, but SD was lower. Inter-individual variability did not justify “extreme position” or “surgical mistakes” in most of the patients for cup inclination.

The hip-spine relationship should be constantly taken into consideration when positioning THA implants to produce optimal clinical results [16]. Several parameters need to be evaluated, and various preoperative planning methods have been proposed [20,21,22]. The OPS functional preoperative study (OPS™, Corin Ltd., Cirencester, UK) allows for the concomitant evaluation of several hip and spine parameters using a validated algorithm and patient-specific planning with computer simulation of the implant postoperative result. Surgeons may accept or refuse and modify the planning to plan the “best” implant position and size in terms of stability and ROM. Lastly, 3D-printed patient’s specific intraoperative instrumentation was provided to obtain the proper position of the implant as planned.

Several recent studies evaluated the risk for dislocation according to the different hip/spine parameters, and most concluded that, in high-risk patients, “a more anteverted cup or more constrained implants (e.g., dual mobility cup) should be considered” [20, 23, 24]. However, no study specified how anteverted the cup should be. The OPS system integrates all the mentioned parameters and produces a dynamic simulation of the reciprocal hip-spine movements. Through this functional analysis, the patient’s specific “best” cup position, implant size, height of femoral neck osteotomy, head length, and others (which are not the focus of the current study) for each patient are provided.

To our knowledge, this was the first study to evaluate a large series of patients’ functional preoperative planning to identify the patient’s specific functional cup position, suggesting that a “general safe zone” in hip replacement surgery may have to be contrasted to the patient’s functional safe zone. Furthermore, the system adopted in the current study simultaneously assessed the “femoral side” of the implant, which was often neglected by most of the preoperative imaging evaluation systems.

We have recently reported imaging and clinical results in the same 100 THAs performed using the OPS system [21]. One dislocation occurred 4 months postoperatively. In that patient, a surgical error resulted in higher postoperative anteversion than planned. Furthermore, a postoperative ± 5° difference from the planned values was found in 78% and 81% of patients for inclination and anteversion, respectively, confirming the high accuracy of the OPS system.

The general validity of the LSZ has been recently called into question [7, 13, 15]. The poor accuracy in achieving proper cup orientation during surgery, the true predictive value in preventing THA dislocation, and the functional acetabular mobility during postural changes represent issues of the LSZ [7, 13, 15, 25,26,27,28]. However, most of those studies failed to define a “new safe zone”. Furthermore, many of these studies are based on 2D imaging and/or do not consider the overall hip-spine or the femoral parameters contributing to THA stability [297, 30, 31]. Thus, functional sagittal cup orientation for THA stability is clinically relevant [2, 16, 32, 33].

In a radiographic evaluation of 206 THA dislocations (2% dislocation rate), 58% of them occurred within the LSZ [1]. Hence “the ideal cup position for some patients may lie outside the Lewinnek safe zone” [1], and more sophisticated analyses should be performed to identify the proper cup position for each patient.

Similarly, evaluating a large institutional registry of more than 7000 THAs for a minimum of six postoperative months [2], 147 (2.1%) dislocations were identified, with 54% of them being within the LSZ. Comparison between the patients and a matched group of stable THAs revealed no differences in acetabular inclination or anteversion angles between dislocators and non-dislocators in all the evaluated zones (5° to 40° anteversion and 25° to 55° inclination, with ± 5° boundaries between zones). Therefore, a truly “safe zone” on the acetabular position likely does not exist, and additional research with 3D imaging technology (e.g., CT scans and EOS® imaging) is needed to assess the functional position of the cup based on the hip-spine relationship. Many of the recent studies, that criticized the validity of the LSZ, were based on 2D images and did not consider the 3D functional cup orientation in the different positions of the body. Those studies, however, did not suggest how to identify a patient-specific safe zone and only functional personalized preoperative planning may provide such information.

Tezuca et al. [16] and Heckmann et al. [17] investigated the functional cup position with regard to the sagittal hip-spine motion and assessed them through pre- and postoperative lateral standing and sitting spine-pelvis-hip radiographs. In 86% of the cases, the LSZ matched the sagittal functional safe zone, but 14% of the hips that were inside the LSZ were not within their sagittal functional safe zone. Furthermore, 90% of patients with late dislocation presented an abnormal sagittal hip motion. The main preoperative predictor of being outside the functional safe zone was preoperative femoral motion, followed by a lesser pelvic motion and abnormal pelvic incidence values. Only dynamic studies of the hip may allow for the identification of patients at risk, but it was not possible to clearly define the angles within which the cup should be positioned for a corresponding value of the hip-spine parameters. This, and the lack of an association between the risk of dislocation and cup orientation angles, represent the two main limitations of their study.

In this regard, many studies [1, 16, 20, 23, 24, 26, 32, 34,35,36,37,38,39,40] suggested increasing anteversion and/or ante-inclination of the cup in patients with a progressive degenerative spine with coexisting muscle atrophy, or previous spinal surgery. In such a situation, reduced pelvic mobility (i.e., reduced pelvic tilt from standing to sitting and subsequent lesser increase of ante-inclination in the sitting position) makes them prone to posterior dislocation. However, none of these studies clarified the amount of anteversion needed to compensate for spinal stiffness.

The main limitation of the current study consisted of the lack of a correlation between dislocation and coronal/sagittal cup positions. However, the present study aimed to describe the discrepancy between LSZ and the patient’s specific safe zone and was not a study regarding dislocations following THA. Furthermore, we were not able to provide specific cut-off values for the several hip/spine parameters corresponding to specific cup position angles. These could be topics for future investigations. A further limitation of the study was the number of patients included. However, this was dictated by our workload and by the fact that the investigation was undertaken in a national health service hospital, where only limited funding for the use of the OPS system was available.

Conclusion

In conclusion, the LSZ, which is based on a static coronal position of the acetabulum, is not applicable to all patients, given the recent understanding of the relationship between the lower lumbar spine, the pelvis, and the hip on the sagittal plane. Acetabular inclination and anteversion alter in the different body positions during daily life activities, and these changes are patient-specific. The current study showed that, when a patient’s specific preoperative planning, based on functional sagittal radiographs of the hip and spine in combination with a lower limbs CT, was performed, the LZS did not correspond to the patient’s functional safe zone in about 8% of patients. A greater discrepancy was found for cup anteversion than cup inclination. According to such findings, the concept of a universal safe zone may have to be abandoned and a functional personalized safe zone may have to be considered. Further research is needed to confirm such findings, to assess the relationship between suggested cup position and hip/spine parameters, and to assess the clinical results of the proposed system. Such investigations will require large cohorts, be randomized, involve several surgeons, be multicentered, and be performed over several years.

Data availability

The datasets used during this study are available from the corresponding authors on reasonable request.

References

  1. Abdel MP, von Roth P, Jennings MT, Hanssen AD, Pagnano MW. What safe zone? the vast majority of dislocated thas are within the Lewinnek safe zone for acetabular component position. Clin Orthop Relat Res. 2016. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s11999-015-4432-5.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Esposito CI, Gladnick BP, Lee YY, Lyman S, Wright TM, Mayman DJ, Padgett DE. Cup position alone does not predict risk of dislocation after hip arthroplasty. J Arthroplasty. 2015;30:109–13. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2014.07.009.

    Article  PubMed  Google Scholar 

  3. Hailer NP, Weiss RJ, Stark A, Karrholm J. The risk of revision due to dislocation after total hip arthroplasty depends on surgical approach, femoral head size, sex, and primary diagnosis. An analysis of 78,098 operations in the Swedish Hip Arthroplasty Register. Acta Orthop. 2012;83:442–8. https://doiorg.publicaciones.saludcastillayleon.es/10.3109/17453674.2012.733919.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Mahoney CR, Pellicci PM. Complications in primary total hip arthroplasty: avoidance and management of dislocations. Instr Course Lect. 2003;52:247–55 PMID: 12690853.

    PubMed  Google Scholar 

  5. Padgett DE, Warashina H. The unstable total hip replacement. Clin Orthop Relat Res. 2004;420:72–9. https://doiorg.publicaciones.saludcastillayleon.es/10.4103/0019-5413.39667.

    Article  Google Scholar 

  6. Patel PD, Potts A, Froimson MI. The dislocating hip arthroplasty: prevention and treatment. J Arthroplasty. 2007;22:86–90. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2006.12.111.

    Article  PubMed  Google Scholar 

  7. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–20 PMID: 7142237.

    Article  CAS  PubMed  Google Scholar 

  8. Grammatopoulos G, Thomas GE, Pandit H, et al. The effect of orientation of the acetabular component on outcome following total hip arthroplasty with small diameter hard-on-soft bearings. Bone Joint J. 2015;97-B:164–72. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.97B2.34294.

    Article  CAS  PubMed  Google Scholar 

  9. Schmalzried TP, Guttmann D, Grecula M, Amstutz HC. The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J Bone Joint Surg. 1994;76-A:677–88. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/00004623-199405000-00008.

    Article  Google Scholar 

  10. De Haan R, Campbell PA, Su EP, De Smet KA. Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg. 2008;90-B:1158–63. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.90B9.19891.

    Article  Google Scholar 

  11. Kennedy JG, Rogers WB, Soffe KE, et al. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13:530–4. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/s0883-5403(98)90052-3.

    Article  CAS  PubMed  Google Scholar 

  12. Paterno SA, Lachiewicz PF, Kelley SS. The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg Am. 1993;79:1202–10. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/00004623-199708000-00013.

    Article  Google Scholar 

  13. Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, et al. The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s11999-010-1487-1.

    Article  PubMed  Google Scholar 

  14. DelSole EM, Vigdorchik JM, Schwarzkopf R, Errico TJ, Buckland AJ. Total hip arthroplasty in the spinal deformity population: does degree of sagittal deformity affect rates of safe zone placement, instability, or revision? J Arthroplasty. 2017. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2016.12.039.

    Article  PubMed  Google Scholar 

  15. Sadhu A, Nam D, Coobs BR, Barrack TN, Nunley RM, Barrack RL. Acetabular component position and the risk of dislocation following primary and revision total hip arthroplasty: a matched cohort analysis. J Arthroplasty. 2017. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2016.08.008.

    Article  PubMed  Google Scholar 

  16. Tezuka T, Heckmann ND, Bodner RJ, Dorr LD. Functional safe zone is superior to the lewinnek safe zone for total hip arthroplasty: why the Lewinnek safe zone is not always predictive of stability. J Arthroplasty. 2019;34(1):3–8. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2018.10.034.

    Article  PubMed  Google Scholar 

  17. Heckmann N., McKnight B., Stefl, M., Trasolini, N.A., Ike H., Dorr L.D. Late Dislocation Following Total Hip Arthroplasty: Spinopelvic Imbalance as a Causative Factor. J. Bone Jt. Surg. Am. 2018; 1845–1853. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.18.00078.

  18. DiGioia AM, 3rd, Jaramaz B, Colgan BD. Computer assisted orthopaedic surgery. Image guided and robotic assistive technologies. Clin Orthop Relat Res 1998;Sep(354): p.8–16. PMID: 9755759

  19. Lazennec JY, Brusson A, Rousseau MA. Lumbar-pelvic-femoral balance on sitting and standing lateral radiographs. Orthop Traumatol Surg Res. 2013;99(1 Suppl):87–103. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.otsr.2012.12.003.

    Article  Google Scholar 

  20. Snijders TE, Schlösser TPC, Heckmann ND, Tezuka T, Castelein RM, Stevenson RP, Weinans H, de Gast A, Dorr LD. The effect of functional pelvic tilt on the three-dimensional acetabular cup orientation in total hip arthroplasty dislocations. J Arthroplasty. 2021;36(6):2184–2188.e1. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2020.12.055.

    Article  PubMed  Google Scholar 

  21. L.Previ,E.Viglietta,V.Giuliani,F.Corsetti,A.Redler,A.Speranza,A. De Carli, R.Iorio. Patient-Specific Instrumetation with laser guided Navigated THA:Clinical and CT evaluation of the fisrt 100 cases; Prosthesis 2023; https://doiorg.publicaciones.saludcastillayleon.es/10.3390/prosthesis5040077.

  22. Eftekhary N, Shimmin A, Lazennec JY, Buckland A, Schwarzkopf R, Dorr LD.A systematic approach to the hip-spine relationship and its applications to total hip arthroplasty. Bone Joint J. 2019 https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.101B7.BJJ-2018-1188.R1.

  23. Dorr LD, Callaghan JJ. Death of the Lewinnek “Safe Zone.” J Arthroplasty. 2019. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2018.10.035. PMID: 30527340.

    Article  PubMed  Google Scholar 

  24. Murphy WS, Yun HH, Hayden B, Kowal JH, Murphy SB. The Safe Zone Range for Cup Anteversion Is Narrower Than for Inclination in THA. Clin Orthop Relat Res. 2018;476(2):325–35. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s11999.0000000000000051.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Kanawade V, Dorr LD, Wan Z. Predictability of acetabular component angular change with postural shift from standing to sitting position. J Bone Joint Surg Am. 2014. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.M.00765.

    Article  PubMed  Google Scholar 

  26. Ike H, Dorr LD, Trasolini N, Stefl M, McKnight B, Heckmann N. Spine-pelvis-hip relationship in the functioning of a total hip replacement. J Bone Joint Surg Am. 2018. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.17.00403.

    Article  PubMed  Google Scholar 

  27. Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22(4):815–21. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.orthres.2003.11.001.

    Article  PubMed  Google Scholar 

  28. Elkins JM, Callaghan JJ, Brown TD. The 2014 Frank Stinchfield Award: the “landing zone” for wear and stability in total hip arthroplasty is smaller than we thought: a computational analysis. Clin Orthop Relat Res. 2015;473(2):441–52. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.orthres.2003.11.001.

    Article  PubMed  Google Scholar 

  29. Huppertz A, Radmer S, Asbach P1, et al. Computed tomography for preoperative planning in minimal-invasive total hip arthroplasty: Radiation exposure and cost analysis. Eur J Radiol. 2011;78:406–13. 18. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.ejrad.2009.11.024.

  30. Nomura T, Naito M, Nakamura Y, Ida T, Kuroda D, Kobayashi T, Sakamoto T, Seo H. An analysis of the best method for evaluating anteversion of the acetabular component after total hip replacement on plain radiographs. Bone Joint J. 2014;96:597–603. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.96B.33013.

    Article  PubMed  Google Scholar 

  31. Tannast M, Murphy SB, Langlotz F, Anderson SE, Siebenrock KA. Estimation of pelvic tilt on anteroposterior x rays a comparison of six parameters. Skeletal Radiol. 2006;35:149–55. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00256-005-0050-8.

    Article  CAS  PubMed  Google Scholar 

  32. Lazennec JY, Brusson A, Rousseau M.THA Patients in standing and sitting positions: a prospective evaluation using the low-dose "full-body" eosA® imaging system. Semin Arthroplasty 2012. https://doiorg.publicaciones.saludcastillayleon.es/10.1053/j.sart.2013.01.005.

  33. Pierrepont JW, Feyen H, Miles BP, Young DA, Bare JV, Shimmin AJ. Functional orientation of the acetabular component in ceramic-on-ceramic total hip arthroplasty and its relevance to squeaking. Bone Joint J. 2016. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.98B7.37062.

    Article  PubMed  Google Scholar 

  34. McCarthy TF, Alipit V, Nevelos J, Elmallah RK, Mont MA. Acetabular cup anteversion and inclination in hip range of motion to impingement. J Arthroplasty. 2006;31:264–8. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2016.01.067.

    Article  Google Scholar 

  35. Tiberi JV 3rd, Antoci V, Malchau H, Rubash HE, Freiberg AA, Kwon YM. What is the fate of total hip arthroplasty (THA) acetabular component orientation when evaluated in the standing position? J Arthroplasty. 2015;30:1555–60. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2015.03.025.

    Article  PubMed  Google Scholar 

  36. Timperley AJ, Biau D, Chew D, White House SL. Dislocation after total hip replacement there is no such thing as a safe zone for socket placement with the posterior approach. Hip Int. 2016;26:121–7. https://doiorg.publicaciones.saludcastillayleon.es/10.5301/hipint.5000318.

    Article  PubMed  Google Scholar 

  37. Yoon YS, Hodgson AJ, Tonetti J, Masri BA,Duncan CP.. Resolving inconsistencies in defining the target orientation for the acetabular cup angles in total hip arthroplasty. Clin Biomech (Bristol, Avon). 2008;253–259. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.clinbiomech.2007.10.014.

  38. Esposito C, Miller T, Kim H, Barlow B, Wright T, Padgett D. Does degenerative lumbar spine disease influence femoroacetabular flexion in patients undergoing total hip arthroplasty? Clin Orthop Relat Res. 2016. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s11999-016-4787-2.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Snijders T, Schlosser T, van Stralen M, Castelein RM, Stevenson RP, Weinans H. The effect of postural pelvic dynamics on the three-dimensional orientation of the acetabular cup in total hip arthropasty. Clin Orthop Relat Res. 2020. https://doiorg.publicaciones.saludcastillayleon.es/10.1097/CORR.0000000000001489.

    Article  PubMed Central  Google Scholar 

  40. Heckmann H, Tezuka T, Bodner R, Dorr L. Functional anatomy of the hip joint. J Arthroplasty. 2020. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2020.07.065.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The authors stated that they did not receive any funds, grants, or other support during the preparation of this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study’s conception and design. D.P., A.R., and C.M. contributed to data collection and analysis. E.V. and F.C. contributed to writing the first draft of the manuscript. Y.G. made the graphics and tables. R.I. and N.M. contributed to the supervision of the study. All authors reviewed and edited the previous version of the manuscript. All authors read and approved the final version of the manuscript.

Corresponding author

Correspondence to Nicola Maffulli.

Ethics declarations

Ethics approval and consent to participate

This study was approved by our University ethics committee and conducted in accordance with the ethical standards in the 1964 Declaration of Helsinki.

Consent for publication

All participants provided informed consent for their inclusion in the study, which was approved by the university's ethics committee.

Competing interests

All authors declared that there are no conflicts of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Iorio, R., Viglietta, E., Corsetti, F. et al. Lewinnek zone not “the be-all and end-all” functional planning for acetabular component positioning in total hip arthroplasty. Arthroplasty 7, 2 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s42836-024-00284-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s42836-024-00284-w

Keywords