Handheld imageless robotic total knee arthroplasty improves accuracy and early clinical outcomes when compared with navigation
Arthroplasty volume 7, Article number: 18 (2025)
Abstract
Background
This study compared imageless robotic-assisted total knee arthroplasty (RATKA) with accelerometer-based navigation (ABN) systems in terms of surgical accuracy and early clinical outcomes.
Methods
A retrospective analysis was conducted on 153 patients (178 knees) who had undergone primary TKA from 2017 to 2023. Surgical accuracy and functional outcomes were assessed up to 12 months post-operation using the Chi-square test, Student’s t-test, and ANCOVA. Subgroup analyses based on patient demographics were also conducted.
Results
Among 153 patients, 101 underwent RATKA, and 52 received ABN. RATKA demonstrated superior alignment accuracy with a significantly lower deviation from the planned alignment (P < 0.05). Additionally, RATKA led to significantly better postoperative functional scores at 6 weeks (P = 0.001) and 3 months (P = 0.001), even after adjusting for preoperative functional differences.
Conclusions
RATKA offers enhanced precision and improves early recovery compared to ABN, supporting its potential as a preferred technology for TKA. Its ability to optimize kinematic alignment may contribute to superior patient outcomes. Compared to ABN, RATKA provides a unique advantage by achieving greater accuracy in planned alignment, which may translate into improved functional recovery. Further research with larger cohorts is recommended to confirm these findings.
Introduction
Total knee arthroplasty (TKA) is a widely performed procedure for pain relief and functional restoration in degenerative joint disease, with the long-term survival being longer and global demand on the rise [1, 2]. By 2030, TKA procedures are projected to grow by 85% (1.26 million procedures) globally, with U.S. demand expected to rise by 139% by 2040 and 469% by 2060, while the U.K. demand is estimated to increase by 117% by 2030 [3, 4]. Despite these advancements, achieving optimal alignment remains a significant challenge, as malalignment is associated with polyethylene wear, prosthetic loosening, and increased failure rates [5]. Furthermore, patients with alignment outliers—such as distal femoral angle (DFA), proximal tibial angle (PTA), or posterior slope angle (PSA)—report worse functional outcomes and greater dissatisfaction compared to non-outliers [6]. Despite improvements in surgical techniques and surgeon expertise, alignment outliers persist at high rates, which continue to impact patient satisfaction and long-term implant survival [7].
To enhance alignment accuracy, computer-assisted TKA technologies have been developed to improve the accuracy of the implant position and alignment, categorized into passive, semi-active, and active systems [8]. Passive systems provide intraoperative guidance requiring full surgeon control, while semi-active systems offer real-time feedback and robotic assistance to enhance precision with surgeon control, and active systems autonomously perform bone resections without direct surgeon involvement [9]. Robot-assisted TKA (RATKA) systems, such as NAVIO and CORI, fall under the semi-active category and employ intraoperative registration and robotic guidance to refine bone resections and optimize soft tissue balancing [10]. Accelerometer-based navigation (ABN) systems, such as KneeAlign 2, are passive systems that rely on position sensors to guide bone resections and assess alignment without preoperative imaging [11].
Recent studies have compared RATKA and ABN with conventional techniques for TKA. RATKA has shown improved postoperative alignment and clinical outcomes compared to conventional methods [12]. Similarly, ABN demonstrated higher rates of neutral alignment postoperatively [13, 14]. However, limited research has directly compared their impact on early functional outcomes and surgical accuracy.
This study aimed to address this gap by evaluating and comparing the early functional outcomes and surgical accuracy of semi-active imageless RATKA and passive imageless ABN in primary TKA, seeking to provide insights into their relative effectiveness to optimize TKA techniques and improve patient outcomes.
Materials and methods
Study design
This was a retrospective analysis of prospectively collected data in the authors’ institutional joint registry. Ethical approval was obtained from the Institutional Ethics Review Committee of the Joint CUHK-NTEC Clinical Research Ethics Committee.
Study population
This study included female and male patients over 40 years old who suffered from end-stage osteoarthritis of the knee (Kellgren and Lawrence grade 3–4) and underwent navigation or robotic-assisted primary total knee replacement from 2017 to 2023. Patients were excluded if they had undergone previous surgery on the same knee, including previous knee arthroplasty or osteotomy or had underlying diseases or abnormal anatomy complicating the surgery, including previous periarticular fracture, severe fixed flexion contracture > 20°, multi-ligament instability, bone stock deficiency requiring augmentation and stems, neuromuscular disorder, acute and chronic infections.
Intervention
All surgeries were performed in a tertiary referral hospital and its affiliated joint replacement center by the same team of experienced specialist orthopedic surgeons from the arthroplasty division. The experienced arthroplasty surgeons performed at least 20 TKA with navigation or robotic system before the start of the recruitment. Comparable experience with the use of the two systems was ensured.
Two navigation systems were utilized in the study: Total knee replacement with a Hand-held accelerometer-based navigation (ABN) system KneeAlign 2 (KA2) (OrthAlign, Inc.; Aliso Viejo, CA, USA) and total knee replacement with a NAVIO or CORI robot (Smith and Nephew, USA). The choice of systems was based on the availability of the systems in the hospital at the time of surgery. All TKAs were performed using either kinematic alignment (KA) or mechanical alignment (MA) according to the surgeons’ preference. Surgical techniques were standardized. All TKAs were performed via a standard medial para-patellar approach with measured resection technique, soft tissue balancing, and then implantation of components with antibiotics-loaded cement. Implants used were either Legion or Journey II BCS total knee system (Smith & Nephew, plc.; Watford, UK). In the KA2 group, the system was employed for proximal tibial and distal femoral osteotomy. In the robotic group, mapping was done and the bone cutting was completed using a combination of burrs and saws.
All patients received identical wound closure techniques and postoperative recovery protocol (adult joint reconstruction enhanced recovery after surgery perioperative analgesic & antiemesis protocol). All patients followed the standard physiotherapy adult joint reconstruction rehabilitation protocol. Patients were discharged from the hospital once their mobility allowed for outpatient care.
Radiographic analysis
Preoperative and postoperative anterior–posterior (AP) standing long-leg radiographs were reviewed by two independent reviewers who were blinded to the intervention grouping. Measurements of hip knee ankle angle (HKA) were performed manually using digital measurement tools. HKA measures the angle between the mechanical axis lines of the femur (from the femoral head center to the intercondylar fossa center) and of the tibia (from the tibial interspinous groove to the tibial mid-plafond). The measurements are further detailed in Fig. 1.
The difference between planned alignment and measured alignment (postoperative HKA) is calculated by subtracting the planned angle from the postoperative HKA. A positive difference indicates a valgus angulation of the knee, while a negative difference is indicative of varus angulation. The absolute difference is then obtained by converting the negative values into positive values to obtain the accuracy of alignment.
Power analysis
A G*Power analysis was conducted based on the sample size. An independent t-test was selected with a post hoc power analysis to compute achieved power. Given a two-tailed input, an effect size of 0.5, and an α of 0.05, the calculated power was 0.8933.
Statistical analysis
Demographic data (age, sex, BMI, degree of varus deformity, range of motion, and the side of operation) were compared between the ABN group and the RATKA group using Student’s t-test or Chi-square test. Operative parameters, including operating time, tourniquet time, pressure, and blood loss, were extracted. Knee Society Knee Score (KSS) and Knee Society Function Score (KSFS) at preoperation, 6 weeks, 3 months, 6 months, and 12 months post-operation were compared between the two groups using Student’s t-test. Pre- and postoperative HKA were compared between the two groups using Student’s t-test. ANCOVA was then used to remove the effects of covariates, including preoperative functional scores between patients when comparing postoperative data between the two groups. Further subgroup analysis based on age, BMI, and sex was done to stratify the results. Inter-rater reliability was assessed to compare radiological measurements in terms of two-way random effects model Cronbach α values (Supplementary Information: Table S1). All statistical analyses were carried out using IBM SPSS version 28 (Armonk, NY: IBM Corp). Statistical significance was set at P < 0.05.
Results
A total of 153 patients (178 knees) were recruited into the study. 101 patients (112 knees) were in the robotic group, while 52 patients (66 knees) were in the navigation group. The preoperative patient demographics between the robotic and navigation groups were generally comparable (Table 1). Significant differences between the preoperative functional scores were observed for both the KSFS (P < 0.001) and KSS (P < 0.001). Other operative parameters are also shown (Supplementary Information: Table S2).
The alignment was significantly more accurate in the robotic group than in the navigation group (P < 0.05) when comparing the planned alignment with the postoperative HKA in terms of absolute difference (Table 2). An absolute difference approach was used to evaluate the degree of variance between the planned alignment and postoperative HKA, eliminating the influence of varus or valgus variance. Moreover, the robotic group showed significantly better patient functional outcomes relative to the navigation group in terms of KSFS at 6 weeks (P = 0.001) and 3 months (P < 0.01).
Further analysis was done to account for the preoperative functional differences between the robotic and navigation groups (Table 3). After controlling for the preoperative functional scores, the robotic group still showed significantly better early postoperative functional outcomes in terms of KSFS at 6 weeks (P < 0.001) and 3 months (P = 0.025). Although ANCOVA was used, the significant differences in the preoperative functional scores introduced confounders to the data.
Further subgroup analysis was performed to stratify the patient groups based on age, BMI, and sex (Supplementary Information: Tables S3–S5). 65 or above was selected as the cut-off for the elderly population as widely defined in the locality. 25 or above was taken as the cut-off for obesity as widely defined in Asian populations.
Discussion
The current study demonstrated that RATKA systems attained significantly better alignment accuracy and early postoperative functional outcomes when compared to ABN systems. Although studies have shown that RATKA or ABN systems significantly improved alignment accuracy and functional outcomes when compared to conventional TKA, there is limited evidence based on direct comparisons between RATKA and ABN systems.
Handheld imageless systems offer several advantages over their image-based counterparts. The reduced need for preoperative imaging lowers the utilization rate of hospital computed tomography (CT) scan [17]. This increases efficiency and allows for better patient selection for CT scans, improving patient flow. Furthermore, the elimination of the preoperative CT scan reduces radiation exposure of patients [15]. Importantly, the learning curve of handheld imageless systems has shown to be comparable to that of existing TKA methods [16]. Consequently, it is important to compare different imageless systems to improve patient outcomes.
RATKA systems are likely to have greater accuracy than navigation systems in bone resection, prosthesis positioning, and gap balancing due to their semi-active nature [17]. While the surgeon has overall control over osteotomy and prosthesis positioning, the system limits the range of movements of the surgical instruments according to the surgeon’s surgical plan [18]. Moreover, the robotic system can potentially augment the surgeon's skills by providing real-time visual, tactile, and auditory feedback for more precise execution of the planned cuts [19]. It was demonstrated that 91.6% of bone resections were within ≤ 1 mm of the preoperative plan in RATKA [17]. Another study also found the accuracy of RATKA in bone resection was high, with 99 of 105 bone resections being within 1 mm of the preoperative plan [20]. In contrast, ABN systems can only provide feedback and warning during the operation without restricting the surgeon [18].
Better accuracy in bone resection in RATKA, as shown in the study, allows kinematic alignment to be performed better. Proper KA could minimize the anatomical change of the bone and the impact on ligament balancefor better implant survival and functional outcomes [21, 22]. This highlights the possible importance of accurate bone resection following a preoperative surgical plan for TKAs.
TKA conducted with RATKA systems may also allow for earlier postoperative recovery. Studies have shown that RATKA results in shorter postoperative lengths of stay due to reduced blood loss [23, 24]. Accurate surgical planning with fewer bone cuts and less soft tissue management along with avoiding instrument insertion that reduces microembolus formation, could all contribute to the reduction in blood loss [24,25,26]. Reduced blood loss could thus reduce the need for transfusion and other associated complications, such as infection, venousthrombolism, and mortality [27]. RATKA is further associated with reduced postoperative pain, opiate analgesia requirements in the early postoperative period, and shorter time to discharge [23].
The benefits of RATKA for early functional outcomes were demonstrated in terms of KSFS at 6 weeks and 3 months. Previous studies found that RATKA sped up the early postoperative functional recovery time, allowing for early mobilization of patients [23, 28]. Early mobilization signifies earlier initiation of physical therapy for patients to train muscle strength and flexibility [29]. Starting rehabilitation within 24 h of TKA could result in greater joint range of motion, improved quadriceps and hamstring strength, and better scores for gait and balance [30]. A study showed that the early rehabilitation group had a better Western Ontario and McMaster Universities Osteoarthritis Index than the standard rehabilitation group at each subsequent visit [31].
Apart from the clinical benefits, the cost-effectiveness of RATKA and ABN is another consideration. A study showed that, compared to conventional standard, imageless handheld RATKA (NAVIO) required a 25-min more surgical time and resulted in an additional cost of USD 2600, while the cost was only USD 650 for imageless navigation [32]. While the surgical time is significantly longer for RATKA (117.44) than navigation (113.33 min), there were no differences in the length of stay, 90-day all-cause revisions, and 1-year PROM scores [33]. This suggests that the navigation system may be more cost-effective. Nevertheless, RATKA may be a cost-effective procedure at high-volume hospitals with a volume of more than 49 procedures per year [34]. This indicates that RATKA is potentially more cost-effective at high-volume hospitals while ABN is more beneficial for low-volume hospitals. Studies on direct comparisons of cost-effectiveness between RATKA and ABN are needed to further help identify suitable candidates for RATKA or ABN.
This study has several limitations. Firstly, the retrospective nature of this study limited the control over sampling and the potential effects of confounding factors. The differences in patient demographics introduced significant confounders to the study. Despite using ANCOVA, the power of the study may be impaired. Patellar resurfacing was not performed in all patients, while 16 studies of 1989 knees showed a significant difference in the functional score for ≥ 5 years between resurfacing and non-resurfacing in TKA [35]. The comparability of surgeons’ experience with RATKA and ABN systems should also be considered due to the learning curve. As a single-center study, the population was more homogenous, and the interventions were tested in relatively optimal conditions, while in the “real-world” setting, TKA is performed in patients with clinical and anatomical differences. Along with the small sample size of 178 patients, the external generalizability is therefore limited. Furthermore, limited surgeon experience may also reduce generalizability due to the learning curve of the systems. This study also has low comparability due to the lack of a conventional TKA group to provide a benchmark for evaluating if the differences in functional outcomes can be attributed to the interventions. As a result, the treatment outcomes may be over- or underestimated. Furthermore, this study showed no significant differences in functional outcomes after 3 months, which raises concern about the long-term benefits of RATKA. Therefore, prospective multi-center research comparing conventional TKA with RATKA and ABN with a larger sample size and longer follow-up periods should be conducted to validate the findings and guide clinical decision-making.
Conclusion
In summary, the improved accuracy and early postoperative functional outcomes of RATKA may implicate it as the preferred technology for knee arthroplasties.
Data availability
Data are available on request due to privacy/ethical restrictions.
References
Scott CEH, Bell KR, Ng RT, MacDonald DJ, Patton JT, Burnett R. Excellent 10-year patient-reported outcomes and survival in a single-radius, cruciate-retaining total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2019;27(4):1106–15. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00167-018-5179-9.
Gao J, Xing D, Dong S, Lin J. The primary total knee arthroplasty: a global analysis. J Orthop Surg Res. 2020;15(1):190. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13018-020-01707-5. Published 2020 May 26.
Shichman I, Roof M, Askew N, et al. Projections and epidemiology of primary hip and knee arthroplasty in medicare patients to 2040–2060. JB JS Open Access. 2023;8(1):e22.00112. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.OA.22.00112. Published 2023 Feb 28.
Jabbal M, Burt J, Clarke J, Moran M, Walmsley P, Jenkins PJ. Trends in incidence and average waiting time for arthroplasty from 1998–2021: an observational study of 282,367 patients from the Scottish arthroplasty project. Ann R Coll Surg Engl. 2024;106(3):249–55. https://doiorg.publicaciones.saludcastillayleon.es/10.1308/rcsann.2023.0039.
Hadi M, Barlow T, Ahmed I, Dunbar M, McCulloch P, Griffin D. Does malalignment affect revision rate in total knee replacements: a systematic review of the literature. Springerplus. 2015;4:835. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40064-015-1604-4. Published 2015 Dec 30.
Kazarian GS, Haddad FS, Donaldson MJ, Wignadasan W, Nunley RM, Barrack RL. Implant malalignment may be a risk factor for poor Patient-Reported Outcomes Measures (PROMs) following Total Knee Arthroplasty (TKA). J Arthroplasty. 2022;37(6S):S129–33. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2022.02.087.
Kazarian GS, Lawrie CM, Barrack TN, et al. the impact of surgeon volume and training status on implant alignment in total knee arthroplasty. J Bone Joint Surg Am. 2019;101(19):1713–23. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.18.01205.
Antonios JK, Korber S, Sivasundaram L, et al. Trends in computer navigation and robotic assistance for total knee arthroplasty in the United States: an analysis of patient and hospital factors. Arthroplast Today. 2019;5(1):88–95. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.artd.2019.01.002. Published 2019 Mar 12.
Jacofsky DJ, Allen M. Robotics in arthroplasty: a comprehensive review. J Arthroplasty. 2016;31(10):2353–63. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2016.05.026.
Schrednitzki D, Horn CE, Lampe UA, Halder AM. Imageless robotic-assisted total knee arthroplasty is accurate in vivo: a retrospective study to measure the postoperative bone resection and alignment. Arch Orthop Trauma Surg. 2023;143(6):3471–9. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00402-022-04648-2.
Budhiparama NC, Lumban-Gaol I, Ifran NN, Parratte S, Nelissen R. Does accelerometer-based navigation have any clinical benefit compared with conventional TKA? A systematic review. Clin Orthop Relat Res. 2019;477(9):2017–29. https://doiorg.publicaciones.saludcastillayleon.es/10.1097/CORR.0000000000000660.
Agarwal N, To K, McDonnell S, Khan W. Clinical and Radiological Outcomes in Robotic-Assisted Total Knee Arthroplasty: A Systematic Review and Meta-Analysis. J Arthroplasty. 2020;35(11):3393–409.e2. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2020.03.005.
Treu EA, Frandsen JJ, Woodley CD, Loughmiller SK, Blackburn BE, Peters CL. Accelerometer-Based Navigation in Primary Total Knee Arthroplasty Leads to Improved Alignment but No Change in Patient-Reported Outcomes. J Arthroplasty. 2023;38(6S):S222–6. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2023.02.081.
Luan Y, Wang H, Zhang M, Li J, Zhang N, Liu B, et al. Comparison of navigation systems for total Knee Arthroplasty: a systematic review and meta-analysis. Front Surg. 2023;17:10. https://doiorg.publicaciones.saludcastillayleon.es/10.3389/fsurg.2023.1112147.
Hirschmann MT, Avram G, Graichen H, Tandogan RN, Mengis N, Zaffagnini S. Same same but different-Image-based versus imageless robotic-assisted total knee arthroplasty! J Exp Orthop. 2024;11(4):e70062. https://doiorg.publicaciones.saludcastillayleon.es/10.1002/jeo2.70062. Published 2024 Oct 19.
Savov P, Tuecking LR, Windhagen H, et al. Imageless robotic handpiece-assisted total knee arthroplasty: a learning curve analysis of surgical time and alignment accuracy. Arch Orthop Trauma Surg. 2021;141:2119–28. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00402-021-04036-2.
Li C, Zhang Z, Wang G, et al. Accuracies of bone resection, implant position, and limb alignment in robotic-arm-assisted total knee arthroplasty: a prospective single-centre study. J Orthop Surg Res. 2022;17(1):61. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13018-022-02957-1. Published 2022 Jan 29.
Yang P, He R, Lei K, Liu L, Yang L, Guo L. Clinical evaluation of the first semi-active total knee arthroplasty assisting robot made in China: a retrospective propensity score-matched cohort study. Int J Surg. 2023;109(6):1552–60. https://doiorg.publicaciones.saludcastillayleon.es/10.1097/JS9.0000000000000322. Published 2023 Jun 1.
Chen X, Deng S, Sun ML, He R. Robotic arm-assisted arthroplasty: the latest developments. Chin J Traumatol. 2022;25(3):125–31. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.cjtee.2021.09.001.
Sires JD, Craik JD, Wilson CJ. Accuracy of bone resection in MAKO total knee robotic-assisted surgery. J Knee Surg. 2021;34(7):745–8. https://doiorg.publicaciones.saludcastillayleon.es/10.1055/s-0039-1700570.
Howell SM, Shelton TJ, Hull ML. Implant survival and function ten years after kinematically aligned total knee arthroplasty. J Arthroplasty. 2018;33(12):3678–84. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2018.07.020.
Hsu CE, Huang JT, Tong KM, Huang KC. Total knee arthroplasty according to the original knee phenotypes with kinematic alignment surgical technique-early clinical and functional outcomes. BMC Musculoskelet Disord. 2020;21(1):839. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s12891-020-03862-6. Published 2020 Dec 11.
Kayani B, Konan S, Tahmassebi J, Pietrzak JRT, Haddad FS. Robotic-arm assisted total knee arthroplasty is associated with improved early functional recovery and reduced time to hospital discharge compared with conventional jig-based total knee arthroplasty: a prospective cohort study. Bone Joint J. 2018;100-B(7):930–7. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.100B7.BJJ-2017-1449.R1.
Lee JH, Jung HJ, Choi BS, Ro DH, Kim JI. Effectiveness of robotic arm-assisted total knee arthroplasty on transfusion rate in staged bilateral surgery. J Clin Med. 2023;12(14):4570. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/jcm12144570. Published 2023 Jul 9.
Kalairajah Y, Cossey AJ, Verrall GM, Ludbrook G, Spriggins AJ. Are systemic emboli reduced in computer-assisted knee surgery?: A prospective, randomised, clinical trial [published correction appears in J Bone Joint Surg Br. 2006 Oct;88(10):1407]. J Bone Joint Surg Br. 2006;88(2):198–202. https://doiorg.publicaciones.saludcastillayleon.es/10.1302/0301-620X.88B2.1690.
Khan H, Dhillon K, Mahapatra P, et al. Blood loss and transfusion risk in robotic-assisted knee arthroplasty: a retrospective analysis. Int J Med Robot. 2021;17(6):e2308. https://doiorg.publicaciones.saludcastillayleon.es/10.1002/rcs.2308.
Hart A, Khalil JA, Carli A, Huk O, Zukor D, Antoniou J. Blood transfusion in primary total hip and knee arthroplasty. Incidence, risk factors, and thirty-day complication rates. J Bone Joint Surg Am. 2014;96(23):1945–51. https://doiorg.publicaciones.saludcastillayleon.es/10.2106/JBJS.N.00077.
Wainwright TW, Gill M, McDonald DA, et al. Consensus statement for perioperative care in total hip replacement and total knee replacement surgery: Enhanced Recovery After Surgery (ERAS®) Society recommendations. Acta Orthop. 2020;91(1):3–19. https://doiorg.publicaciones.saludcastillayleon.es/10.1080/17453674.2019.1683790.
Jette DU, Hunter SJ, Burkett L, et al. Physical therapist management of total knee arthroplasty. Phys Ther. 2020;100(9):1603–31. https://doiorg.publicaciones.saludcastillayleon.es/10.1093/ptj/pzaa099.
Labraca NS, Castro-Sánchez AM, Matarán-Peñarrocha GA, Arroyo-Morales M, Sánchez-Joya Mdel M, Moreno-Lorenzo C. Benefits of starting rehabilitation within 24 hours of primary total knee arthroplasty: randomized clinical trial. Clin Rehabil. 2011;25(6):557–66. https://doiorg.publicaciones.saludcastillayleon.es/10.1177/0269215510393759.
den Hertog A, Gliesche K, Timm J, Mühlbauer B, Zebrowski S. Pathway-controlled fast-track rehabilitation after total knee arthroplasty: a randomized prospective clinical study evaluating the recovery pattern, drug consumption, and length of stay. Arch Orthop Trauma Surg. 2012;132(8):1153–63. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/s00402-012-1528-1.
Christen B, Tanner L, Ettinger M, Bonnin MP, Koch PP, Calliess T. Comparative cost analysis of four different computer-assisted technologies to implant a total knee arthroplasty over conventional instrumentation. J Personal Med. 2022;12(2):184. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/jpm12020184.
Singh V, Fiedler B, Simcox T, Aggarwal VK, Schwarzkopf R, Meftah M. Does the Use of Intraoperative Technology Yield Superior Patient Outcomes Following Total Knee Arthroplasty?. J Arthroplasty. 2021;36(7S):S227–32. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.arth.2020.11.001.
Hua Y, Salcedo J. Cost-effectiveness analysis of robotic-arm assisted total knee arthroplasty. Huang KC, ed. PLoS One. 2022;17(11):e0277980. https://doiorg.publicaciones.saludcastillayleon.es/10.1371/journal.pone.0277980.
Chen K, Dai X, Li L, Chen Z, Cui H, Lv S. Patellar resurfacing versus nonresurfacing in total knee arthroplasty: an updated meta-analysis of randomized controlled trials. J Orthop Surg Res. 2021;16(1):83. https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13018-020-02185-5.
Acknowledgements
Not applicable.
Funding
The authors received no financial support for the research, authorship, and/or publication of this article.
Author information
Authors and Affiliations
Contributions
The authors confirm contribution to the paper as follows: study conception and design: J.T., M.O.; data collection: J.T., A.T., C.W., G.L., T.C., R.M., J.N., K.H., M.O., P.Y.; analysis and interpretation of results: J.T., M.O.; manuscript preparation: J.T., A.T., M.O. All authors reviewed the results and approved the final version of the manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Ethical approval was obtained from the Institutional Ethics Review Committee of the Joint CUHK-NTEC Clinical Research Ethics Committee.
Consent for publication
Not applicable.
Competing interests
M.T.Y.O. is the Editorial Board of Arthroplasty. The authors declare that they have no conflicts of interest.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
42836_2025_303_MOESM1_ESM.docx
Supplementary Material 1. Table S1. Cronbach α values, Table S2. Operative Parameters, Table S3. Subgroup Analysis Age – Student’s T-Test.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Tran, J.YS., Tang, A.YT., Wong, CK. et al. Handheld imageless robotic total knee arthroplasty improves accuracy and early clinical outcomes when compared with navigation. Arthroplasty 7, 18 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s42836-025-00303-4
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s42836-025-00303-4